Prma 460 29 User Manual
Description
3.12.5 Parameter area
The parameter area (1) dsplays parameters that
are applcable to the selected mode of operaton
Parameters are greyed out f not applcable
Parameter Adult Value Paedatrc value
Set Tdal Volume 600 ml 150 ml
Set mnute volume 6 ltres 22 ltres
T+PS Int 10 cmHO 10 cmHO
Set BPM 10 15
Set IE rato 120 120
PEEP OFF OFF
Set Pressure lmt 38 cmHO 38 cmHO
Set trgger 08 cmHO 08 cmHO
Apnoea Alarm Trgger
Mute lmt s adustable n
Spontaneous mode only
15 s 15 s
Volume type Tdal Tdal
Parameters dsplayed at Start-up
VT SET
The parameter s shown as ether Tdal Volume or Mnute Volume
dependent on user set menu optons
Ventlator mode Set volume dsplay type
STANDBY / PV / SPONT / VV / SIMV / PSV Tdal or Mnute
SMMV Mnute
Unts
Mnute volume Ltres
Tdal volume Dependent on the current volume settng
For tdal volume n the range 40 to 990 ml the unts are
mllltres, wth a resoluton of 10 ml
For tdal volumes above 990 ml, the unts are Ltres
It s not possble to set a tdal or mnute volume value that wll
result n
a) The nspraton tme beng less than 200 ms, or
b) The requred drve gas flow fallng outsde the range 2 to
75 L/mn
Attempts to do ths wll result n an “INORRET RATE OR
RATIO” alarm beng dsplayed
1