4.10 Oxygen Monitor (Fuel cell sensor)
NOTE
If a Masmo Nomolne ISA AX+ or OR+ Module s ftted, the ntegral oxygen montor system (fuel cell sensor type) s dsabled
Measurement Range 0 - 100%
Resoluton ± 1%
Accuracy and Lnearty ± 2% of full scale (at constant temperature and pressure)
Response Tme 90% of fnal value n approxmately 15 seconds (ar to 100% O)
Operatng Temperature -20 to 50° (-4 to 112°F)
Storage Temperature -5 to 40° (23 to 104°F)
Relatve Humdty Range 0 to 99% (non-condensng)
Hgh Prorty Alarm Flashng, 2 x 5 audo pulses wth 6 seconds repeat tme
Medum Prorty Alarm Flashng, 3 audo pulses wth 24 seconds repeat tme
Low Prorty Alarm Statc wth sngle beep sound
Alarm Mute 30 seconds for hgh prorty alarm
120 seconds for medum prorty alarm
Low Alarm Set Range 18% - 99% (± 1%)
Hgh Alarm Set Range 19% - 105% (± 1%)
Sensor
Type alvanc fuel cell sensor (0 - 100%)
Lfe 1500000 O% hours at 20° (68°F)
800000 O% hours at 40° (104°F)
Interference ases and Vapours
Interference Test as Mxture Interference n O%
Ntrous Oxde 80% NO + 20% O 1 to 15%
arbon Doxde 5% O + 665% NO + 285% O <1%
Helum 50% He + 50% O <1%
Halothane 4% Halothane + 672% NO + 288% O 15 to 2%
Enflurane 5% Enflurane + 665% NO + 285% O 12 to 18%
Isoflurane 5% Isoflurane + 665% NO + 285% O 12 to 18%
Sevoflurane 5% Sevoflurane + 665% NO + 285% O 1 to 15%
Humdty Effects
Sensor output s relatvely unaffected by prolonged operaton n ether hgh or very low relatve humdty
If the sensor shows sgns of beng affected by condensaton, dry the sensor wth soft tssue
AUTION
DO NOT use heat to dry the sensor
Temperature Effects
The sensor has a bult-n temperature compensaton crcut, and s relatvely unaffected by temperature changes wthn the
operatng temperature range gven above
Pressure Effects
The sensor measures O partal pressure, and ts output wll rse and fall due to pressure change (eg changes n barometrc
pressure, or breathng system pressure)
An ncrease n pressure of 10% at the sensor nlet wll produce a 10% ncrease n sensor output
NOTE
Alttude compensaton s automatcally appled durng calbraton
Specification