Prma 460 84 User Manual
Operating the Anaesthesia System
7.3 Ventilator Default Settings and Start-up
Functon Ste Default (range or opton) Adult Default Paedatrc Default
VT (mL) 20 - 1600 600 150
VM (L/mn) 05 - 50 60 22
T+PS INIT (cmHO) 4 - 70 10 10
Set BPM 4 - 100 10 15
I  E 102 to 18 12 12
PEEP (cmHO) Off 4 - 20 (opton Off 4 - 30) OFF OFF
Lmt (cmHO) 38 38 38
Trgger (L/mn) 1 1 1
Apnoea alarm mute lmt (s)
(selectable n Spontaneous
mode only)
15, 30, 60, 120, 180 15 15
Volume type Tdal or Mnute Tdal Tdal
Back lght level (%) 0 - 100 100 100
7.3.1 Default Settings Selection
At start-up, the ntroducton screen allows the user to select
one of three default settngs and vew the parameters, as
shown above
ADULT DEFAULTS
PAEDIATRI DEFAULTS
SITE DEFAULTS
The optons wll reman, even after the ventlator s turned
off
NOTE
1 The user must select one of the above default groups
before the ventlator wll swtch to standby n that default
mode
2 SITE DEFAULT s edtable n standby mode
a) Adust the parameter values from wthn the Servce
menu (SITE DEFAULTS)
b) Press to confrm the new settngs for ste defaults