Pre-use Checks
6.6 System Alarm
T
e
s
t
WARNIN
The anaesthetc machne must not be used f any alarm,
montorng, or protecton system devces are not functonng
correctly
Prmary Oxygen Falure
A
l
a
r
m
The machne s ftted wth a warnng whstle and a vsual
ndcator (1)
These components act as oxygen supply falure devces and
consttute the prmary alarm system, powered only by the
resdual oxygen supply, as descrbed n secton 343
The system can be checked whenever the low pressure
oxygen system s frst pressursed by turnng on a cylnder or
connectng a ppelne
1 The whstle wll sound brefly as pressure ncreases, and
2 The vsual ndcator (1) wll turn from red to green
Whstle, Vsual
Indcator,
and as ut-off De
v
c
e
T
e
s
t
A formal test (ncludng the acton of the nternal gas cut-off
devce) s performed as follows
1 onnect oxygen, ntrous oxde and ar supples
2 Set the system swtch (2) to ON
3 Set a flow of 2 L/mn on all gases
4 Dsconnect the oxygen supply at the wall socket or close
the oxygen cylnder valve and check
a) That as the oxygen flow slows down, the whstle starts
to sound and contnues for at least 7 seconds
b) That the flow of ntrous oxde s cut off completely
before the oxygen flowmeter shows zero flow
c) That the vsual ndcator (1) turns red before the
oxygen flow s entrely stopped
d) Ar, f ftted, contnues to flow
e) Renstate the oxygen supply
heck that the flow of ntrous oxde s renstated, and
that the vsual ndcator (1) turns green agan
NOTE
All gases must be ncluded n the pre-use check
1
2