EasyManua.ls Logo

Penlon Prima 460 - Auxiliary Gas Outlets - Optional; Oxygen Failure Warning Devices; Oxygen Flush; Fresh Gas Delivery

Penlon Prima 460
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Prma 460 56 User Manual
Specification
4.3 Auxiliary Gas Outlets - Optional
Oxygen
One self-sealng connecton on rear of machne
Ar (on machnes wth Ar supply opton)
One self-sealng connecton on rear of machne
Supply pressure
Ppelne supply n use as s suppled at ppelne supply
pressure
ylnder supply as s suppled at reduced
pressure from the cylnder supply
secondary regulator
Flow rate
Flow rate for each gas 60 L/mn maxmum
WARNIN
Flow rates greater than 60 L/mn could affect the fresh gas
flow to the patent
4.4 Oxygen Failure Warning Devices
1 as system whstle
2 Vsual ndcator, drect pressure operated
4.5 Oxygen Flush
Button on the front edge of worksurface
The system supples 25-75 L/mn when the button s fully
depressed
4.6 Fresh Gas Delivery
Mechancal AHD System
Mnmum oxygen concentraton 30% ± 3% (of total O +
NO flow)
Oxygen basal flow 50-75 ml/mn
Maxmum Flow
Oxygen 12 L/mn
Ntrous oxde 12 L/mn
Ar 12 L/mn
4.7 Electrical supply
Power nput 100-240 VA, 50-60 Hz, 1000-2400 VA maxmum, 10 A
Power cable Permanently attached lead (3 m)
Power Outlets Three outlets, country specfc
Fuse (mans supply) Type T 2 AH, fused L and N
2 A, 250 V ratng, 5x20 mm, ant-surge, ceramc
Fuse (battery) Type T 5 AH,
5 A, 250 V ratng, 5x20 mm, ant-surge, ceramc

Table of Contents