Description
Default settng
To adust the default settng for the apnoea alarm mute, use
the SITE DEFAULT menu opton
Apnoea alarm trgger
The apnoea alarm wll trgger f the patent fals to breathe
after 15 seconds Ths settng cannot be changed
Apnoea message dsplay
Durng the alarm mute perod the apnoea alarm message
wll be dsplayed f the alarm has been trggered The alarm
message wll be removed f the patent breathes
If re-trggered, the alarm message wll be dsplayed agan, but
the audble alarm wll reman muted f the mute perod s stll
actve