Prma 460 60 User Manual
Specification
Pressure lmt (volume control mode) 10 to 80 cmHO
Fresh as ompensaton Automatc tdal volume adustment
Inspratory Flow 2 to 75 L/mn
Inspratory Tme 03 – 10 seconds (normal)
03 – 5 seconds (effectve n support modes)
Expratory Tme 03 – 10 seconds
(effectve dependent on Inspratory tme)
Modes Off
Standby
Volume ycle
Pressure ontrolled
Spontaneous
Volume ycle
Sgh Sgh can be set from 1n, where n has a range of 10 to 100
Inspratory Pause Inspratory pause can be vared n the menu from 0 - 60%
Pressure range 5 - 70 cmHO (pressure control mode)
Electronc PEEP 4 - 20 cmHO (or optonal 4 - 30 cmHO)
Spontaneous Mode Actve Volume and Pressure alarms
Sprometry Resoluton ±10 ml
Ventlator Accuracy
NOTE
1 Ambent temperature of 20° (±10%) and ambent atmosphere of 1013 kPa (±10%))
2. Accuracy with Spirometry malfunctioning may vary from the figures given below.
Delvered volumes ±10% (>300 ml)
PRV ±15% (>300 ml)
±20% (100 to 300 ml)
±50%, or ± 10 ml, whchever s greater (<100 ml)
Hgh arway pressure lmt ±5 cmHO
Rate (bpm) ±10%
Tdal volume + (50±5%)