OPERATION MULTI GAUGE CONTROLLER
3.6.11.2 DISPLAY SETTINGS
To change display settings go to: Setup Menu →Device Settings →Display
In Display submenu it is possible to change the following parameters:
Brightness - display brightness value. Its value can be changed in the range of 10-100%.
Touch Screen Autolock - When the value is set to ON, the autolock function is active. If the touch
panel is not used for longer than 3 min then the screen is locked. 3 minutes - screen is locked.
Unlock the device by pressing ”Yes” in the displayed message.
Customer Name - The device can be assigned an individual name that appears on the top of main
window. This allows to distinguish between several devices of the same type.
System Date - The device has a built-in real time clock (RTC)(RTC). The menu allows to set the
current date and time.
When setting a date separate the individual parts with the sign “.”. Example: 21-03-2014
When setting a time separate the individual parts with the sign “:”. Example: 13:20:22. Alterna-
tively, the following time format is accepted: 3:4:5 -> 03:04:05, 12:8:1 -> 12:08:01
3.6.11.3 LANGUAGE MENU
This submenu allows to change the device language.
In order to select the language go to Setup menu →Device Settings →Language.
To complete changing of the language the device must be restarted.
3.6.11.4 CHART ACTIVATION
Setup menu →Device Settings →Chart Activation
The submenu is used to activate the chart module. If the module is active this submenu is not
available. For details see chapter 3.4.2.7.
3.6.12 SAVING/LOADING SETTINGS
All relevant parameters of the device may be saved automatically in a non-volatile memory and
restored after turning the power back on. In addition, the current coniguration of the device may
be saved at any moment for restoration at a later time. Up to 6 different operation conigurations of
the device can be saved, and each coniguration can be assigned an individual name. It is possible
to return to the saved coniguration or to the default settings at any time.
In order to turn on or turn off the autosave function of parameters, one should:
1. Go to menu Setup Menu →Save/Load .
2. Click on Autosave in order to change its status.
To save the coniguration of the device:
1. Enter Setup Menu →Save/Load →Save.
2. Choose a slot from the list where the settings will be saved (previous parameters will be
deleted) and click on it.
3. Enter the required name using the keyboard.
4. Click Enter to conirm the name.
Page 72 User manual rev. 03