EasyManuals Logo
Home>Rigol>Test Equipment>MSO2000A Series

Rigol MSO2000A Series Programming Guide

Rigol MSO2000A Series
376 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #322 background imageLoading...
Page #322 background image
RIGOL 2 Command System
2-306 MSO2000A/DS2000A Programming Guide
:TRIGger:SHOLd:TYPe
Syntax
:TRIGger:SHOLd:TYPe <type>
:TRIGger:SHOLd:TYPe?
Description
Set or query the hold type of setup/hold trigger.
Parameter
Name Type Range Default
<type> Discrete {SETup|HOLd|SETHOLd} SETup
Explanation
SETup: set the time (refer to the
:TRIGger:SHOLd:STIMe command) that the data stays stable and
constant before the clock edge appears.
HOLd: set the time (refer to the
:TRIGger:SHOLd:HTIMe command) that the data stays stable and
constant after the clock edge appears.
SETHOLd: set the time (refer to the
:TRIGger:SHOLd:STIMe and :TRIGger:SHOLd:HTIMe
commands) that the data stays stable and constant before and after the clock edge appears.
Return Format
The query returns SET, HOL or SETHOL.

Table of Contents

Other manuals for Rigol MSO2000A Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO2000A Series and is the answer not in the manual?

Rigol MSO2000A Series Specifications

General IconGeneral
BrandRigol
ModelMSO2000A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals