R&S®SML / R&S®SMV03 Stereo Modulation (Option R&S SML-B5)
1090.3123.12 4.35 E-6
PTY
Description: Sets or reads the program type (PTY).
Command: PTY=xx
Query: PTY?
Response: xx
Value range: 00 to 31 (ASCII coded decimal numbers)
Example: Command:
STEReo:DIRect “PTY=08“
Sets the program type to be transmitted to "08".
Query:
STEReo:DIRect? “PTY“
Response:
"08"
Note: A two-digit value must always be set. A leading zero, if any, must also be specified.
PTYN
Description: Sets or reads the program type (PTY) name.
Command: PTYN= xxxxxxxx
Query: PTYN?
Response: xxxxxxxx
Value range: 8 ASCII-Zeichen
Example: Command:
STEReo:DIRect “PTYN=Football“
Sets the program type name to be transmitted to "Football".
STEReo:DIRect “GS=0A,10A”
Group 10A is activated in addition to group 0A. The program type name "Football"
is now transmitted.
Query:
STEReo:DIRect? “PTYN“
Response:
"Football"
Note: An eight-digit value must always be set. Blank spaces, if any, must also be entered,
otherwise the value will not be accepted.