13 Trouble Shooting
13.1 Possible problems
Page 104
CT8000P-2
Art. no.: 2.511407 Rev.: a
No printout of
interpretation
statement,
averaged cycles or
measurements
Incorrect settings
Check that the interpretation and measurement options are ena-
bled for the printout and that the lead sequence is set to Normal
(see page 75 section 10.3 and page 40 section 4.12.2,
Function keys
blocked
Software hangs up
Function keys defective
Switch off and on again after a few seconds.
Press and hold the On/Off button for 10 seconds to force the device
to switch off. Reconnect mains and switch on.
If the problem persists, call your local SCHILLER representative.
Interferences, lines
on the display
Excessive EMC interferences
Check for sources of excessive EMC interferences.
Memory full
The ECG recording cannot be
stored because the memory is
full.
Delete old ECG recordings, see page 61.