Page 55
Art. no.: 2.511407 Rev.: a
Exercise ECG 7
User guide Safety notes 7.1
CT8000P-2
7 Exercise ECG
7.1 Safety notes
The CARDIOVIT AT-102 G2 is CF classified. The patient connection is fully iso-
lated. Always ensure however, that during the recording neither the patient nor
the conducting parts of the patient connector nor the electrodes come into con-
tact with other persons or conduction objects (even if these are earthed).
Do not use the unit or the ergo device if the earth connection is suspect or if the
mains cable is damaged in any way.
Before starting an exercise ECG, make sure the ergometer's user guide is read
and understood. The instructions given in this user guide do not override those
for the ergometer.
Ensure that the resting ECG is normal and that the patient is physically fit
enough to carry out an exercise ECG.
Ensure a charged defibrillator is to hand when carrying out an exercise test.
To avoid possible interference from the ergometer when carrying out an exercise
test, it is recommended that both the CARDIOVIT AT-102 G2 and the ergometer
are connected to the same common ground.
The potential equalisation connector is situated on the rear of the unit. A yellow/
green ground cable is supplied as an option (Article number 2. 310 005).