Page 95
Art. no.: 2.511407 Rev.: a
Maintenance 12
User guide Maintenance interval table 12.1
CT8000P-2
12 Maintenance
12.1 Maintenance interval table
The regular system maintenance must include a software check according to the
manufacturer's instructions. The test results must be recorded and compared to the
values in the accompanying documents.
Maintenance work not described in this section may only be performed by a qualified
technician authorised by seca.
The following table indicates the intervals and responsibilities of the maintenance
work required. Local regulations in your country may stipulate additional or different
inspection intervals and tests.
Interval Maintenance step Responsible
Before each use • Visual inspection of the device and ECG electrodes User
Every 6 months
• Visual inspection of the device
(see page 101, 12.7 Inspection Report)
– Function key test
– Keyboard test
– Cables and accessories
– Mains cable
• Functional tests according to the instructions (see page 101,
12.7 Inspection Report)
User
Every 12 months • Safety test according to IEC/EN 62353
Qualified service
personnel