2 Introduction
2.6 Connections
Page 18
CT8000P-2
Art. no.: 2.511407 Rev.: a
2.6 Connections
2.6.1 Back panel
(1) Power supply unit connection 100 -220 VAC, 50/60 Hz, 1.3 A.
(2) Potential equalisation stud. The potential equalisation stud is used to equalise
the ground potential of the unit to that of any nearby mains-powered equipment.
Use the hospital or building common ground for all mains-powered units.
(3) RJ-45 Ethernet LAN connection (Local Area Network)
(4) 2 USB interfaces for the barcode scanner and USB memory sticks.
(5) 2x RS-232-DB9 (COM1 connections for ergometer, COM2 not supported)
(6) ECG patient cable connector DB15
(7) Kensington lock
All externally connected hardware must be approved by seca. Connection of any
hardware not approved by seca is at the owner‘s risk. Moreover, the unit's
warranty may become invalid.
2
1
3
4
5
6
7
The patient cable as well as the connector (6) comply with the safety standard CF
, i.e. they are fully floating and isolated and defibrillation protected.
The unit is only CF rated and defibrillation protected if used with the original
SCHILER patient cable.