14 Technical Data
14.2 ECG
Page 108
CT8000P-2
Art. no.: 2.511407 Rev.: a
14.2 ECG
Interfaces • ECG cable interface
• Potential equalisation
• Network connection (1Gbit)
•2 USB
• 2 RS-232
Memory
Memory for at least 350 ECG recordings, 100 resting rhythm recordings and 10
exercise ECGs.
Patient input • Fully floating and isolated, defibrillation-protected (only with original SCHILLER
patient cable)
Lead configuration • Standard 12-lead
• Right precordials
• Standard C4r
• Balanced
• Left Posterior
•Nehb
• Paediatric
Display
Leads
Status
• 6- to 12-channel display of the selected leads
– Paper speed of 12.5/ 25/ 50 mm/s
– Amplitude of 5 /10 / 20 mm/mV
• Filter status
• Power source
• Leads
• Electrode contact status
• Heart rate (HR)
• Date and time
• Patient name and number
• LAN/WLAN transmission status
Filter
Myogram filter (muscle tremor)
Notch filter
• Set to 25, 40, 150, 250 Hz (250 Hz = Filter Off)
• Distortion-free suppression of superimposed AC 50 or AC 60 Hz sinusoidal inter-
ferences by means of adaptive digital filtering
Data record
With optional interpretation ETM
• Patient data
• Listing of all ECG recording data (date, time, filter)
• ECG measurements results (intervals, amplitudes, electrical axes)
• Averaged complexes
• Guidance on interpreting adult and paediatric ECGs
ECG amplifier
Complies with IEC 60601-2-25 and ANSI/AAMI EC11