Page 89
Art. no.: 2.511407 Rev.: a
General and System Settings 10
User guide Exercise ECG 10.8
CT8000P-2
10.8.3 Bike protocol
10.8.4 Treadmill protocol
Pre-defined bike protocols.
25/25-2/25 corresponds to a warm-up load of 25 Watt, a base load of 25 Watt, an
increase of 25 Watt per step for 2 minutes as well as a recuperation load of 25 Watt.
Menu Parameter Description
Bike protocol
Protocols:
• 25/25-2/25
• 30/30-3/25
• 30/40-3/25
• 50/25-2/25
• 50/50-3/25
• 75/25-2/25
• Conconi
Order and settings options for exercise ECG display:
Select (with key OK or Activate) and definition of the order
(keys Up/Down) of the following protocols.
Use ramp
Yes/No. Activating the ramp in steps of 1 Watt from one step to the
next.
Use warm-up
Yes/No. Activating warm-up phase after start of the exercise ECG
as first step of the exercise protocol.
Menu Parameter Description
Treadmill protocol
Protocols:
• Mode-Bruce
• Balke-Ware
•Bruce
• Cornell
• Ellestad
• Mod-Balke
• Mod-Balke-Ware
• Mod-Naughton
• Slow USAFSAM
• USASAM
Order and settings options for exercise ECG display:
Select (with key OK or Activate) and definition of the order
(keys Up/Down) of the following protocols.
Use ramp
Yes/No. Activating the ramp in steps of 1 Watt from one step to the
next.
Use warm-up
Yes/No. Activating warm-up phase after start of the exercise ECG
as first step of the exercise protocol.