158 FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
Commissioning
Subject to change without notice
5.7.2.4 Enabling (starting) DataLogs
To enable (start) a disabled DataLog, complete the steps described under
pg. 158, 5.7.2.5
(Resetting DataLog Parameters to Defaults).
5.7.2.5 Resetting DataLog parameters to defaults
Complete the following steps to reset the DataLog parameters to the defaults:
Go to the DataLogs page (select Meter / DataLogs from menu).
Choose the "Configuration" tab (see
Figure 76).
Switch the meter into Configuration Mode (choose "File / Configuration Mode" from the
menu).
Click "Reset defaults".
Before resetting DataLog Parameters to Defaults
In MEPAFLOW600 CBM from V1.1.00, DataLog entries are not saved in the
meter database. Before clearing entries from DataLogs, download and export
the entries into Excel (
pg. 178, 6.4.2).
Defaults
The default settings are described in
2.8.3.1 and
2.8.3.2.