EasyManua.ls Logo

SICK microScan3 Core I/O AIDA

SICK microScan3 Core I/O AIDA
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.11.2 Settings for monitoring cases
Name
Ent
er a name which is as descriptive as possible for the monitoring case in the Name
field. If you create a lot of monitoring cases, you should consider a naming concept that
makes it possible to identify the monitoring cases easily (for example right cornering,
left cornering).
7.11.3 Cut-off paths
You can create cut-off paths and define the outputs switched by the cut-off paths.
Y
ou need a cut-off path for every field in a field set. If the field sets have different sizes,
use the field set with the most fields as a guide.
Creating cut-off path and entering a name
1.
Create a cut-off path for every field in the largest field set.
2. Enter a descriptive name for each cut-off path.
Assigning an OSSD pair
b
Place a check in the box for the OSSD pair.
The OSSD pair is assigned to the cut-off path.
7.11.4 Assigning field sets
Assigning a field set to a monitoring case
The field sets that have been created are listed in the Field sets area.
1. Create cut-off paths, see "Cut-off paths", page 93.
2. Drag the field set onto the monitoring case.
The fields in a field set are arranged as they were drawn in the field editor.
Deleting the assignment of a field set from the monitoring case
b
Dr
ag the field set from the monitoring case table onto the trash-can icon.
The field set is removed from the relevant monitoring case.
Defined cut-off behavior
DANGER
H
azard due to lack of effectiveness of the protective device
Persons and parts of the body to be protected may not be recognized in case of
non-observance.
The Always ON (safe) function has the same effect as a field which is always clear. In
a monitoring case with the Always ON (safe) function, the cut-off path containing this
function is permanently in the ON state.
b
Dr
ag the Always OFF function onto the cut-off path.
The field is viewed as being permanently interrupted.
The cut-off path is always in the OFF state.
b
Drag the Always ON (safe) function onto the cut-off path.
The field is viewed as being permanently clear.
The cut-off path is always in the ON state.
If fields have not been assigned to certain cells in a monitoring case table, Safety
D
esigner assigns the Always OFF function to these cells.
CONFIGURATION 7
8017784/1ELL/2022-01-21 | SICK O P E R A T I N G I N S T R U C T I O N S | microScan3 Core I/O AIDA
93
Subject to change without notice

Table of Contents

Other manuals for SICK microScan3 Core I/O AIDA

Related product manuals