CHAPTER 7 Quality control
7-6 Sysmex XE-5000 Instructions for Use
Revised April 2007
3. The QC Selection screen displays a list of the QC files.
Select the file by Lot No., mode and level into which you
wish to save analysis results and press “Select” on the
function menu.
4. Analyze control blood with reference to “Chapter 6: 6.17
Sample analysis in manual mode”.
QC analysis in sampler mode
“Execute L-J” can be performed in sampler mode if either a
specified control sample is used or barcode labels (for QC file
use) are affixed on a sample tube.
1. Prepare control blood.
2. Affix the barcode label to the correct position on the
sample tube.
3. Analyze control blood with reference to “Chapter 6: 6.19
Sample analysis in sampler mode”.
After analysis is completed, the results are recorded
automatically in the file designated by the barcode.
QC analysis in manual closed mode
QC analysis can be performed in closed mode if control blood
is used.
1. From the Main Menu screen on the Main Unit, select
“QC” on the function menu.
The QC Menu screen will appear.
2. From the QC Menu screen, select “Exec. QC” on the
function menu.
The QC Selection screen will appear.
3. The QC Selection screen displays a list of the QC files.
Select the file by Lot No., mode and level into which you
wish to save analysis results and press “Select” on the
function menu.
4. Analyze control blood with reference to “Chapter 6: 6.20
Sample analysis in manual closed mode”.
Note:
For instructions on how to affix barcode labels, see
“Chapter 6: 6.16 Preparing sample analysis”.