CHAPTER 8 Calibration
8-4 Sysmex XE-5000 Instructions for Use
Revised April 2007
Exclusion
When the compensation rate is far away from 100% because
of insufficient mixing or analysis error, the data can be
excluded from calibration calculation. If necessary, the
excluded data can be restored.
Exclusion
1. Using the and keys, move the cursor to select the
data to exclude.
2. Select “Exclude” on the function menu of the Automatic
Calibration screen.
The data directed by the cursor will be crossed off, and
the cursor will move to the next line. The averages of the
reference value, analysis value and compensation rate
will be calculated again and displayed.
Canceling the exclusion
1. Using the and keys, move the cursor to select the
data whose calibration value you wish to recalculate.
2. Select “Exclude” on the function menu of the Automatic
Calibration screen.
The line on the data will disappear, and the cursor will
move to the next line. The averages of the reference
value, analysis value and compensation rate will be
calculated again and displayed.
Executing the calibration
Update the compensation rate to the new compensation rate
calculated from the averages of the reference value and
analysis value.
1. Select “Execute” on the function menu of the Automatic
Calibration screen.
The calibration will be executed, and the new
compensation rate will appear in the “New” column.
Execute Exclude Last Data
Next No.123456789012346
DP No. 123456789012345
Manual
8
9
10
AVG
15.9
16.0
15.2
15.3
47.0
47.1
47.5
46.0
15.6
16.6
15.6
15.4
46.5
47.5
47.5
46.3
101.9
96.4
97.4
99.2
101.1
99.2
100.0
99.4
C
Ready
<Auto Calibration>
Reference
HGB
HGB
HCT
99.2%
99.4%
98.1%
100.0%
HCT HGB HCT HGB HCT
Analyze
[Current] [New]
Comp.[%]
Num
DP
Xm