CHAPTER 9 Cleaning and Maintenance
9-30 Sysmex XE-5000 Instructions for Use
Revised April 2007 EU
5. Remove spout kit fixture No. 3 from the reagent container.
6. Fit spout kit fixture No. 3 onto the new reagent container.
Lift up the aspiration port of the reagent container and
engage it on spout kit fixture No. 3.
Push it all the way in, to avoid the reagent container from
collapsing.
7. Prior to connecting the quick connector to the SIM
cubitainer spout, wipe the inside area of the quick
connector and the SIM cubitainer spout with a lint-free
gauze moistened with ethyl-alcohol. This cleaning will
prevent contamination and lubricate the o-ring when
connecting.
8. Pull up the collar of the quick connector.
Spout kit fixture No. 3