CHAPTER 2 Safety Information
2-8 Sysmex XE-5000 Instructions for Use
Revised April 2007
(4)
*1: After installing, remove the anti-sticking spacers.
(5)
(6)
(7)
(8)
(9)
CAUTION
When installing this instrument, remove the constant-pressure screw, then
take out the anti-sticking spacers which are inserted between the sample rotor
valves. At that stage, rinse the surface of each valve cleanly before setting
them in place.
The fluid channel inside the instruments are rinsed with diluent before
shipping from the factory. Background count may be high immediately after
unpacking, so rinse thoroughly.
RISK OF INFECTION
In principle, all parts and surfaces of the
instrument must be regarded as infective.
Always keep the switch in the analysis position.
WARNING
Never touch the detector when the power
of the Main Unit is turned ON. Otherwise,
electrical shock may result.
WARNING
Do not remove this cover when the power
to the Main Unit is ON. Doing so may
result in injury.
WARNING
Do not put your fingers inside when the
power to the Main Unit is ON. Doing so
may result in injury.