CHAPTER 11 Technical Information
11-8 Sysmex XE-5000 Instructions for Use
Revised April 2007
Stability of blood over
time after sampling
36 hours later
48 hours later
NEUT% Within ±8 NEUT%
LYMPH% Within ±7 LYMPH%
MONO% Within ±3 MONO%
EO% Within ±3 EO%
BASO% Within ±1 BASO%
NEUT% Within ±8 NEUT%
LYMPH% Within ±7 LYMPH%
MONO% Within ±4 MONO%
EO% Within ±3 EO%
BASO% Within ±1 BASO%
Stability of blood over
time after sampling
24 hours later
IG% Within ±1.5 IG%
RET-He Within ±8% (RET#≥0.010 × 10
6
/µL)
IPF Variation Within 30% or Variation range Within 2.0 IPF (PLT 100.0 (× 10
3
/µL) or
more and IPF is 2(%) or more.)
Aspirated blood volume Sampler mode: Approx. 200 µL
Manual Closed mode: Approx. 200 µL
Manual mode: Approx. 130 µL
Capillary mode: Approx. 40 µL (the blood volume required for dilution)
Number of stored
samples
Analysis data with histograms: 10,000 samples
Scattergrams: 10,000 samples
Patient information: For 5,000 patients
Order information: 1,000 samples
Quality control X
control (L-J control): 300 points × 20 files, 51 parameters
X
M control: 300 points × 1 file, 54 parameters
Storage conditions for
transportation
Ambient temperature:-10 – 60ºC
Relative humidity: 95% or less (no condensation)
Information
Use samples at 18–26°C, or samples that have been kept refrigerated
(2–8
°C). Samples are at room temperature or refrigerated, and
refrigerated samples are returned to room temperature before analysis.
Some samples may not enter the above ranges, depending on the state of
preservation etc.
Information
Use samples at 18–26ºC, or samples that have been kept refrigerated
(4–8ºC). Samples are at room temperature or refrigerated, and
refrigerated samples are returned to room temperature before analysis.
Some samples may not enter the above ranges, depending on the state of
preservation etc.
Information
Samples should be stored at room temperature.