CHAPTER 6 Operation
Sysmex XE-5000 Instructions for Use 6-7
Revised July 2007 EU
Samples Analyzed per Reagent Container
• The number of cycles per container refers to the number of
cycles that can be run in CBC+DIFF+RET+NRBC analysis
mode.
Replacing the Reagent
After replacement of the reagent, perform automatic cleaning,
check that the background count is low, then start the sample
analysis.
Reagent Abbreviation
No. of cycles/
container
Capacity
per
container
CELLPACK EPK Approx. 660
cycles
20.0 L
CELLSHEATH ESE Approx. 7300
cycles
20.0 L
STROMATOLYSER-
FB
FBA Approx. 2750
cycles
5 L
STROMATOLYSER-
4DL
FFD Approx. 2750
cycles
5 L
STROMATOLYSER-
4DS
FFS Approx. 2000
cycles
42 mL
STROMATOLYSER-
NR(L)
SNR Approx. 2000
cycles
3.6 L
STROMATOLYSER-
NR(S)
Approx. 2000
cycles
43 mL
SULFOLYSER SLS Approx. 10000
cycles
5 L
STROMATOLYSER-
IM
SIM Approx. 3200
cycles
10 L
RET SEARCH (II)
(Diluent)
RED Approx. 550
cycles
1 L
RET SEARCH (II)
(Dye solution)
Approx. 550
cycles
12 mL
Caution!
• Use reagent that has been left at room temperature
(15–30°C) for at least 24 hours.
• In handling a reagent that may have frozen, follow
the precautions given on the package.
Otherwise analysis may not be performed correctly.
• When replacing the reagent container, make sure
that no dust adheres to the cubitainer spout kit.
Otherwise analysis may not be performed correctly.
• After opening a reagent container, make sure that no
dust, dirt or bacteria comes into the container.
Otherwise analysis may not be performed correctly.