CHAPTER 6 Operation
Sysmex XE-5000 Instructions for Use 6-37
Revised July 2007
14. Once analysis is complete, set Sample to Normal on the
Sample No. Setting screen to return to normal mode.
Display analysis results
The analysis results of Body fluid analysis mode are displayed
on the main menu screen of the Main Unit LCD screen. The
results are also displayed on the sample explorer and data
browser screens (the Body Fluid tab) of the Information
Processing Unit (IPU).
Analysis samples analyzed in body fluid analysis mode are
marked with an “F” mark to the left of the sample No. on the
Main Unit LCD screen.
“F” is displayed in the analysis mode display column of the
sample explorer and data browser screens.
Caution!
If either of the analysis values from body fluid analysis
(WBC-BF and RBC-BF) is high, there is the possibility
of carryover on the analysis results for the next
sample. If analysis results for WBC-BF and RBC-BF
exceed the values in the table on the left, the “Execute
Background check” error is displayed, asking operater
to perform a background check before analyzing the
next sample. Analyze the next sample after performing
the background check by pressing the “B-Check”
function menu on the Main Unit.
Make sure to correct the “Execute Background check”
error before analyzing next sample.
Information
• Only analyze body fluid in body fluid analysis mode.
Always use normal mode for analyzing human blood.
• Body fluid analysis is not compatible with negative
judgment.
• Body fluid analysis results are not subject to X
m
control.
WBC-BF 1.000 [×10
3
/µL]
RBC-BF 1.000 [×10
6
/µL]
Next No. 19750112
DP No.
RBC-BF 0.000 x10^6/uL
WBC-BF 0.000 x10^3/uL
MN 0.000 x10^3/uL ---.-%
PMN 0.000 x10^3/uL ---.-%
C D
FNo. 19750111
Num
DP
Xm
BF Manual
Ready
B-Check AutoRinse Maint Reagent