125
TROUBLESHOOTING
Error occurred during network
setup.
Check settings or connections.
Shared Folder1 :
Shared Folder2 :
Shared Folder3 :
Displayed when any of the shared folder settings fail.
Errors that occurred for each folder are displayed. Please check the network
settings and connection.
For details of the errors, check the description of each error.
Error occurred during network
output.
Check settings or connections.
After checking, output again.
Shared Folder1 :
Shared Folder2 :
Shared Folder3 :
Displayed when data output to any of the shared folder fails.
Errors that occurred for each folder are displayed. Please check the network
settings and connection. Then output again. For details of the errors, please
check the description of each error.
Error occurred during network
output.
Check settings or connections.
After checking, output again.
Shared Folder1 :
Shared Folder2 :
Shared Folder3 :
Rebuild: Rebuild network con-
nection
Close: Check the network set-
tings
Displayed when data output to any of the shared folder fails.
Errors that occurred for each folder are displayed. Please check the network
settings and connection. Then output again. For details of the errors, please
check the description of each error.
Touching the Rebuild button will attempt to rebuild the network. However, it will
not output again.
Alignment operation stopped for
patient’s safety.
Check positional relations
between patient’s eyes and mea-
suring head.
(E17001)
Alignment was stopped because the measuring head was too close to the eye.
Check the positional relations between the patient's measuring eyes and the
measuring head.
Measuring operation stopped for
patient’s safety.
Check positional relations
between patient’s eyes and mea-
suring head.
(E17002)
Measuring operation was stopped because the measuring head was too close
to the eye. Check the positional relations between the patient's measuring
eyes and the measuring head.
Measuring nozzle glass stain
detected.
Refer to manual for cleaning
instructions.
(E17101)
Displayed when a blot is detected on the measuring nozzle and the glass
inside the measuring nozzle in T/P measurement mode. Clean the measuring
nozzle and the glass inside the measuring nozzle by referring to “
CLEANING
THE MEASURING NOZZLE AND THE GLASS INSIDE THE MEASURING
NOZZLE” on page 117.
The battery for the clock may be
low power.
Check the time setting.
(E17103)
Displayed when the battery for the built-in clock becomes run down.
- When the battery consumed, confirm the difference in time and adjust it.
- When the battery becomes completely consumed, to verify whether time
stopping occurred and call your service personnel.
Error [ aaaa : bbbb cccc-dddd ]
Reboot by turning off/on power.
(E17106)
Displayed when a part that is not operating normally is detected by the opera-
tion check inside the main unit. Contact your service personnel.
A query in progress.
Please wait until complete.
Information is being retrieved from the server based on the entered search cri-
teria. Please wait.
Not find the query location.
Check the network connection.
(E18002)
Displayed when the server is not found or when communication has timed out.
Check the connection settings and attempt intercommunication check. After
successful intercommunication, execute QUERY again.