126
TROUBLESHOOTING
Patient ID isn't selected.
Select Patient ID in query result.
Displayed when a patient is not selected on the search result screen. Select a
patient for measurement and press the OK button.
Success intercommunication
check with MWL server.
Displayed upon successful intercommunication check with the server.
Failure intercommunication
check with MWL server.
(E18005)
Displayed upon failed intercommunication check with the server.
ALL modified user settings suc-
cessfully deleted.
Return device to factory settings?
Reset the settings of the main unit to return to the factory default settings.
To execute the process, press the OK button.
Device returned to factory set-
tings.
OK to turn off and on unit.
Displayed when the factory reset is complete.
Please turn the power of the main unit off and then on again for the setting to
take effect.
Do you want to execute it? You can execute the function. To execute the function, press the OK button.
Execution processing failed.
(E91024)
Function execution failed. Check the execution conditions for the function.
Executing.
Please wait for a moment.
Function is being executed. Please wait for the completion of execution.
OVER-SPH Spherical power exceeds -25D or +22D.
Measurement cannot be performed for out of measuring range.
OVER-CYL Cylindrical power exceeds ±10D.
Measurement cannot be performed for out of measuring range.
NO TARGET Displayed when there is no target or the eye image is too dark.
You should tell the patient to open their eyes as wide as possible, and tell not
to move the eyes as possible. Then perform the measurement again. Even if
you cannot perform the measurement after above manner, it may be possible
to measure by changing to the cataract mode(CAT).
ALIGN ERR Displayed when the alignment is significantly failed during the measurement.
You should tell the patient to open their eyes as wide as possible, and tell not
to move the eyes as possible. Then perform the measurement again.
AGAIN Displayed when there is more than ±5D difference from the previous measure-
ment value.
You should tell the patient to open their eyes as wide as possible, and tell not
to move the eyes as possible. Then perform the measurement again.
NO CENTER Displayed when center of eye can not be found.
You should tell the patient to open their eyes as wide as possible, and tell not
to move the eyes as possible. Then perform the measurement again.
ERROR Displayed when the patient blinks or otherwise moves his/her eye during mea-
surement. If this message appears while correctly measuring the standard
accessory model eye in REF/KRT, an instrument malfunction may have
occurred. Contact your service personnel.
OVER-R Corneal curvature exceeds 5.00-10.00mm.
Measurement cannot be performed for out of measuring range.
OVER Displayed when IOP value exceeds 60 mmHg.