Vulnerability and software update information is available at the following link.
https://topconhealthcare.com/product-updates/
The manufacturer or your dealer will provide you with information about the end of security support for the
device.
Software updates include security updates such as SOUPs, etc. The user shall apply the latest software.
If there is a possibility that some security incident has occurred, disconnect the device from the hospital network
and take initial action according to hospital policy, such as running anti-virus software and checking access logs,
if necessary.
User shall implement password authentication and other access control for servers on the network, and disable
legacy versions of the SMB communication protocol. User shall enable SMB protocol encryption for the shared
folders and TLS1.2 or other encryption for SQL server TCP/IP communication.
Electromagnetic Compatibility (EMC)
This instrument has been tested (with 100/120/230V) and found to comply with IEC 60601-1-2:2014+AMD1:
2020 (Ed.4.1). This instrument radiates radio frequency energy within standard and may affect other devices in
the vicinity. If you have discovered that turning on/off the instrument affects other devices, we recommend you
change its position, keep a proper distance from other devices, or plug it into a different outlet. Please consult
your authorized dealer if you have any additional questions.
The external devices to be connected must comply with the corresponding EMC standards(e.g. CISPR 32/CISPR 35).
The patient may be injured. It may affect other devices in the vicinity.