When switching the left and right eyes by the control lever, be sure to return the measuring head to the backward
side by tilting the control lever to the backward before moving the measuring head to the left or right.
It causes injury by hitting the patient's eyes or nose.
Do not use or apply any aerosol-type cleaner near the instrument.
If a drop of cleaner remains inside the measuring nozzle, the patient's eye may be injured during measurement.
To clean the intraocular pressure measuring window glass, measuring nozzle and the glass inside the measuring
nozzle, use ethanol. Using other chemicals may cause damage to the patient's eye during measurement.
Do not use in environments where device that generates strong magnetic fields, such as MRI device is installed.
Preventing Electric Shocks
To avoid electric shocks, do not insert metal objects into the instrument body through the vent holes or gaps.
To avoid electric shocks, do not handle the power plug with wet fingers.
The power cord in standard accessories for this instrument cannot use besides this instrument.
To avoid electric shocks, do not touch the external connection terminal and the patient at the same time.
When connecting this instrument to a commercially available personal computer, use a personal computer that
conforms to IEC 62368-1, and use an insulating device that conforms to IEC 60601-1 in the patient environment.
Do not connect an additional power strip or an extension cord to the system.
The total 1kVA is the maximum allowable load of the auxiliary power supply socket for the insulation transformer,
which is provided for the system.
Do not connect the device exceeding this capacity.
Use the auxiliary power supply socket of the insulation transformer to power only a device that will be a compo-
nent of the system.
It is dangerous to connect any device which is not used as a component of the system, to the insulation trans-
former.
When the insulation transformer is not used, the personal computer and the monitor for the personal computer
must be installed out of the patient environment.
Use of controls or adjustments or performance of procedures other than those specified herein may result in haz-
ardous radiation exposure.
Do not remove the enclosures. Laser high-power is radiated.
Ensuring Security
When connecting this instrument to an external device through LAN, apply the security update to the external
device, make use of anti-virus software and take other countermeasures against computer virus properly.
When connecting this instrument to an external device through LAN, set the ID and password of the user to the
external device.
When outputting data to the shared folder on an external device from this instrument, set a proper user ID and
password to the shared folder.
Do not connect anything other than the specified external devices to the external I/O terminals. Information such
as patient ID may be leaked.
The entered patient ID will be displayed on the control panel. Do not leave the device with the control panel dis-
played for long periods of time. Please handle printouts included patient ID with care. The patient ID information
may be leaked.
When using this instrument in an environment where a third person can see it, turn the control panel so that it
cannot be seen by a third person. Please handle printouts included patient ID with care so that it cannot be seen
by a third person. The patient ID information may be leaked.
When connected with an IT network, ensure the appropriate and sufficient security to prevent the infection with
malware and a computer virus, the leak of information, etc.
There is a risk of data leakage.
Do not connect any USB storage device that is not checked with the anti-virus software to the USB port of this
instrument.
To avoid leakage of personal information, erase the data before discarding the storage device.
If you need information about SBOM, ask your dealer. Information can be provided in xlsx and SPDX-Lite formats.