97
OPTIONAL OPERATIONS
5 When the red or green Z alignment bars disappear and the alignment mark is turned green, the
automatic measurement is performed or press the MEASUREMENT switch.
* When “Auto Shoot” in “Function(Common)2” on the Setup screen is set to ON, operate the control
lever to automatically start measurement once the alignment is complete. Refer to “Auto Shoot” on
page 105.
6 Measurement is performed and measurement values are displayed on the control panel.
NOTE
• Even if fine alignment has not been achieved, measurement can be per-
formed by pressing the MEASUREMENT switch. To ensure correct mea-
surement, try to get fine alignment.
• It cannot be measured again for 1.5 seconds after the measurement.
• Even if you press and hold the MEASUREMENT switch of the control
lever in the manual mode of TONO/ PACHY, it will not measure repeatedly
(only once).