Troubleshooting
Alert
Causes
Recommended
Action
Vibration—In
Test
mode
only
Tip
loose
due
to
incorrect
assembly
Testing
before
sterilization:
1.
Use
the
tip
torquing
base
and
wrench
to
tighten
the
tip.
2.
Test
the
tip.
3.
Sterilize
the
handpiece.
Testing
after
sterilization
when
tip
assembled
in
a
nonsterile
area:
1.
Remove
the
handpiece
from
the
sterile
tield.
2.
Use
the
tip
torquing
base
and
wrench
to
tighten
the
tip.
3.
Test
the
tip.
4.
Sterilize
the
handpiece.
Testing
after
sterile
tip
assembled
to
handpiece
on
the
sterile
field:
1.
Use
the
sterilizable
torque
base
and
a
disposable
sterile
wrench
to
tighten
the
tip.
(Refer
to
the
Supplement,
Assembling
or
Changing
Tips
on
the
Sterile
Field,
which
is
part
2
of
the
CUSA
EXcel
System
User's
Guide.)
2.
Test
the
tip.
Tip
contacting
another
object
Remove
the
contact
with
any
other
object.
4
10-4
CUSA
EXcel-8
System
Service
Manual