Troubleshooting
Alert
Causes
Recommended
Action
Vibration—In
Test
mode
only—
continued
Blocked
tip
1.
Verify
that the
manifold
suction
tube
connects
properly
to
the
nosecone
by
disconnecting,
then
reconnecting,
the
tubing
and
its
clip.
Use
the
tip
cleaner
to
distodge
the
blockage
in
the
tip.
Continue
to
push
the
tip
cleaner
into
the
tip
until
the
cleaner
becomes
visible
in
the
tubing.
When
you
have
cleared
the
blockage,
verify
again
that the
manifold
suction
tube
connects
properly
to
the
nosecone.
If
the
tip
remains
blocked,
disconnect
the
manifoid
suction
tube
connector
from
the
nosecone.
Use
the
tip
cleaner
to
dislodge
blockage
from
the
tubing,
then
reconnect
the
tubing
to
the
nosecone.
Immerse
the
handpiece
tip
in
sterile
irrigation
fluid
briefly.
This
may
clear
the
suction
tubing.
бипооузеапол.
Damaged
or
cracked
tip
Testing
before
sterilization:
1.
2.
3.
Use
the
tip
torquing
base
and
wrench
to
replace
the
tip.
Test
the
tip.
Sterilize
the
handpiece.
Testing
after
sterilization
when
tip
assembled
in
a
nonsterile
area:
1.
3.
4.
Remove
the
handpiece
from
the
sterile
field.
Use
the
tip
torquing
base
and
wrench
to
replace
the
tip.
Test
the
tip.
Sterilize
the
handpiece.
Testing
after
sterile
tip
assembled
to
handpiece
on
the
sterile
field:
1.
2.
Use
the
sterilizable
torque
base
and
a
disposable
sterile
wrench
to
replace
the
tip.
(Refer
to
the
Supplement,
Assembling
or
Changing
Tips
on
the
Sterile
Field,
which
is
part
2
of
the
CUSA
EXcel
System
User's
Guide.)
Test
the
tip.
.
GUSA
EXcel-8
System
Service
Manual
105