User
Troubleshooting
Condition
Causes
Recommended
Action
Low
suction
at
the
handpiece
tip—continued
A
blockage
in
the
suction
tubing
between
the
handpiece
and
the
suction
canister
1.
Examine
the
suction
tubing
along
its
entire
length
for
a
blockage.
2.
Squeeze
the
blocked
area
to
loosen
İt.
3.
Immerse
the
handpiece
tip
in
sterile
irrigation
fluid
briefly.
This
may
clear
the
blockage.
Suction
canister
full
Replace
the
suction
canister.
Suction
connections
at
the
suction
canister
not
seated
properly
1.
Remove
any
unused
suction
canister
“elbow”
connectors.
2.
Reseat
all
connections.
Open
port(s)
on
the
suction
canister
Close
any
open
ports
on
the
suction
canister.
Suction
canister
improperly
connected
Refer
to
the
canister
manufacturer's
instructions
for
connecting
the
canister.
Inside
of
the
contamination
guard
wet,
blocked,
or
both
1.
Disconnect
the
contamination
guard
fitting
from
the
suction
canister
tubing.
If
you
feel
no
suction
at
the
end
of
the
contamination
guard
fitting,
the
guard
is
blocked.
2.
Replace
the
contamination
guard.
Problem
at
the
CUSA
EXcel
console
suction
port
Remove
the
contamination
guard
and
put
a
finger
in
the
suction
system
port.
If
you
feel
no
suction,
call
Valleylab
for
assistance.
Suction
pinch
valve
working
improperly
Check
the
Lap
mode
indicator.
e
If
Lap
mode
is
off,
in
Run
mode
the
pinch
valve
should
be
open.
*
Lap
mode
is
on,
in
Run
mode
the
pinch
valve
should
be
open
only
when
you
activate
vibration.
these
conditions
do
not
exist,
call
Valleylab
for
assistance.
The
handpiece
gets
hot
Surgeon
holding
handpiece
at
tip
and
flue
Hold
the
handpiece
at
the
nosecone.
Handpiece
damage
due
to
tip
torquing
without
the
torque
base
Replace
the
handpiece.
The
nosecone
is
loose
Large
o-ring
not
installed
correctly,
or
missing
Install
the
large
o-ring
in
the
proper
position
on
the
handpiece.
Nosecone
worn
Replace
the
nosecone.