User
Troubleshooting
Condition
Causes
Recommended
Action
Low
suction
at
the
handpiece
tip
Low
suction
setting
on
the
control
panel
Increase
the
setting.
System
inadvertently
in
Lap
mode
Turn
off
the
Lap
mode
by
pressing
the
Lap
mode
button.
Suction
tubing
improperly
connected
to
handpiece
Verify
that
the
manifold
suction
tube
connects
properly
to
the
nosecone
by
disconnecting,
then
reconnecting,
the
tubing
and
its
clip.
Blocked
tip
1.
Verify
that the
manifold
suction
tube
connects
properly
to
the
nosecone
by
disconnecting,
then
reconnecting,
the
tubing
and
its
clip.
2.
Use
the
tip
cleaner
to
dislodge
the
blockage
in
the
tip.
Continue
to
push
the
tip
cleaner
into
the
tip
until
the
cleaner
becomes
visible
in
the
tubing.
3.
When
you
have
cleared
the
blockage,
verify
that
the
manifold
suction
tube
connects
properly
to
the
nosecone.
Éunoouss|qnoil
4.
ifthe
suction
remains
low,
disconnect
the
manifold
suction
tube
connector
from
the
nosecone.
Use
the
tip
cleaner
to
dislodge
blockage
from
the
tubing,
then
reconnect
the
tubing
to
the
nosecone.
5.
Immerse
the
handpiece
tip
in
sterile
irrigation
fluid
briefly.
This
may
clear
the
blockage.
Blockage
in
the
suction
tubing
at
the
handpiece
1.
Immerse
the
handpiece
tip
in
sterile
irrigation
fluid
briefly.
This
may
clear
the
suction
tubing.
2.
Press
the
Status
button
to
change
to
Standby
mode.
Use
the
sterile
tip
cleaner
to
clean
out
the
tip,
then
put
the
tip
in
irrigating
fluid
to
flush
the
tubing.
3.
As
necessary,
disconnect,
clean,
and
reconnect
the
suction
tubing
connector
at
the
handpiece.
A
pinch
or
kink
in
the
suction
tubing
Straighten
the
tubing
to
remove
a
kink,
or
remove
anything
that
might
pinch
the
tube.
CUSA
EXcel-8
System
Service
Manual
10-13