R Series Service Manual
1-14 9650-0903-01 Rev. L
4.13 Observe the voltage. Verify the charger voltage is 11.97V-12.43V.
oo
4.14 On the Test Fixture set the switch to 20 Ohms. Verify the charger voltage is 9.50V-11.87V.
oo
4.15 On the Test Fixture set the switch to 27 Ohms. Verify the charger voltage is 11.97V-12.43V.
oo
4.16 Remove the charger load Test Fixture.
4.17 Connect the universal cable to the Impulse
4000.
4.18 Press the ENERGY SELECT UP ARROW
button until 200J is displayed.
Verify 200J is displayed.
oo
4.19 Press the CHARGE button and start timing with
a stopwatch. Stop timing when the SHOCK
button illuminates.
Verify charge time is between 3-10 seconds.
oo
4.20 Press the ENERGY DOWN ARROW button. Verify the unit internally discharged.
oo
Record your results on the Maintenance Test Checklist.
Do this... Observe this... Pass/Fail

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB

Related product manuals