EasyManuals Logo

ZOLL R Series Service Manual

ZOLL R Series
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #309 background imageLoading...
Page #309 background image
Transferring Device Check and Activity Log Files Through Wi-Fi (R Series Plus Only)
9650-0904-01 Rev. N ZOLL R Series Operator’s Guide 12–9
TRANSFER FAILED
Invalid Wi-Fi configuration
Error Number: N
Contact Network Administrator
(cont.)
N = 8: The Defib Upload Server's Username /
Password are blank within the Full Disclosure and Defib
History server configuration items Data COMM II only)
Action:
If Full Disclosure Transfer Mode is set to Server, make
sure that the Username (Full Disclosure) and Password
(Full Disclosure) entries are valid.
If Defib History Transfer Mode is set to Manual,
Automatic, or Both, make sure that the Username
(Defib History) and Password (Defib History) entries are
valid.
N = 9: Missing required certificate file for Network
Profile 1 with Enterprise Authentication selected
(Data COMM II only)
Action:
If Authentication Protocol is set to PEAP, make sure
that a root certificate has been added (if not using the
default ZOLL root certificate).
If Authentication Protocol set to TLS, make sure that a
root certificate has been added (if not using the default
ZOLL root certificate) and that the correct client
certificate has been configured.
N = 10: Missing required certificate file for Network
Profile 2 with Enterprise Authentication selected
(Data COMM II only)
Action: Same as above.
TRANSFER FAILED
Network Error: General
Error Number: NNNN
Contact Network Administrator
The transfer failed because of a general network Wi-Fi
error. NNNN may indicate one of the following error
codes:
N = 7004
N = 7010
N = 7011
N = 7015
N = 7016
Action: Install another ZOLL Data COMM card.
N = 7500
N = 7501
N = 7502
N = 7503
N = 7504
N = 7505
N = 7506
N = 7507
Action: Install another Data COMM II card.
N = 7017
Action: Make sure that the Defib Upload Server is
running, is operational, and can be reached on the
network.

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB