CHAPTER 12 FILE TRANSFER
12–10 www.zoll.com 9650-0904-01 Rev. N
TRANSFER FAILED
Network Error: Unknown [-7002]
Action: Make sure that no illegal characters are used
for the SSID names.
TRANSFER FAILED
Wi-Fi Card Failure
Error Number: -7009
Contact Network Administrator
The Wi-Fi card’s configuration data is corrupt or cannot
be read. (Data COMM only)
Action:
Install another Data COMM card. Contact the Network
Administrator for assistance.
TRANSFER FAILED
Wi-Fi Network Not Found
Contact Network Administrator
The R Series could not associate with a Wi-Fi network.
Action:
1) Make sure that the configuration values for the SSID
names are correct.
2) Make sure that the R Series unit is within range of
the wireless server.
3) Contact the Network Administrator for assistance.
TRANSFER FAILED
Could Not Connect To Server
Contact Network Administrator
The R Series could not communicate with the
DefibUpload Server.
Action:
1) Make sure that the DefibUpload Server is running, is
operational, and can be reached on the network.
2) If WEP is enabled, make sure that the WEP keys on
the R Series and Access Point match (value and slot
location).
TRANSFER FAILED
Network Error: DHCP
Contact Network Administrator
The DHCP request has failed.
Action:
1) If WEP is enabled, make sure that the WEP keys on
the R Series and Access Point match (value and slot
location).
2) Make sure that the DHCP server is operational.
TRANSFER FAILED
Network Error: DNS
Contact Network Administrator
The DNS request has failed.
Action:
1) Make sure that the DNS server is running,
operational, and properly configured.
2) Make sure that the R Series configuration values
pertaining to the server names (FD and DH) are
correct.
TRANSFER FAILED
Full Disclosure Service Not Available
Contact Network Administrator
The option to accept full disclosure files is not enabled
on the server.
Action: Contact the Network Administrator for
assistance.
TRANSFER FAILED
Defib History Service Not Available
Contact Network Administrator
The option to accept Defib History files is not enabled
on the server.
Action: Contact the Network Administrator for
assistance.

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB