Transferring Device Check and Activity Log Files Through Wi-Fi (Optional)
9650-0912-01 Rev. N ZOLL R Series Operator’s Guide 11–9
TRANSFER FAILED
Network Error: General
Errror Number: NNNN
Contact Network Administrator
The transfer failed because of a general network
Wi-Fi error. NNNN may indicate one of the following
error codes:
N = 7004
N = 7010
N = 7011
N = 7015
N = 7016
Action: Install another Data COMM card.
N = 7500
N = 7501
N = 7502
N = 7503
N = 7504
N = 7505
N = 7506
N = 7507
Action: Install another Data COMM II card.
N = 7017
Action: Make sure that the Defib Upload Server is
running, is operational, and can be reached on the
network.
TRANSFER FAILED
Network Error: Unknown [-7002]
Action: Make sure that no illegal characters are used
for the SSID names.
TRANSFER FAILED
Wi-Fi Card Failure
Error Number: -7009
Contact Network Administrator
The Wi-Fi card’s configuration data is corrupt or
cannot be read. (Data COMM only)
Action:
Install another Data COMM card. Contact the Network
Administrator for assistance.
TRANSFER FAILED
Wi-Fi Network Not Found
Contact Network Administrator
The R Series could not associate with a Wi-Fi
network.
Action:
1) Make sure that the configuration values for the
SSID names are correct.
2) Make sure that the R Series unit is within range of
the wireless server.
3) Contact the Network Administrator for assistance.
TRANSFER FAILED
Could Not Connect To Server
Contact Network Administrator
The R Series could not communicate with the
DefibUpload Server.
Action:
1) Make sure that the DefibUpload Server is running,
is operational, and can be reached on the network.
2) If WEP is enabled, make sure that the WEP keys
on the R Series and Access Point match (value and
slot location).
3) Make sure that the server port is correct.

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB