R Series Service Manual
2-2 9650-0903-01 Rev. L
ALARM SET ALARM SET status message when setting alarms.
ANALYSIS HALTED ECG analysis halted due to user interaction such as:
• Lead/size change
Analyze button was pressed again
• Impedance fault
• Charging error detected in auto defib mode
ASYNC PACE ONLY Posted along with other faults. Indicates that the PD
module cannot detect sync pulses.
ATTACH PADS AED: No pads on Auto Defib power-up.
Verify proper OneStep cable/hands-free therapy
electrode connection by disconnecting and
reconnecting the OneStep cable and hands-free
therapy-electrodes.
AUDIO QUEUE FULL Indicates that the audio output queue is full. Additional
voice prompts can't be queued at this time.
None.
BATT HIGH CURRENT Battery is not charged and battery current is greater than
1.6 A.
Replace battery or AC charger.
BATT HIGH VOLTAGE Indicates that the charger voltage is too high. Replace battery or AC charger.
BATT LOW VOLTAGE Indicates that the charger voltage is too low. Replace battery or AC charger.
BATTERY COMMS ERROR Battery is not communicating with the host. Replace battery interconnect board.
BATTERY FAULT Replace battery.
BATTERY ID FAULT Replace battery.
CAL. BARO. PRESSURE Barometric pressure reading is out of range. Calibrate the barometric pressure.
CALIBRATE NIBP NIBP calibration is incomplete or failed. Cycle power and retry; if problem persist calibrate
NIBP system.
CANNOT CHARGE Cannot charge when charge button pressed. Replace high voltage module or capacitor.
Error Message Explanation User
Advisory
Technical Action

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB

Related product manuals