R Series Service Manual
9650-0903-01 Rev. L 2-11
REPEAT NIBP MEAS
The unit exceeded the maximum number of inflation
attempts.
The unit exceeded the 180-second measurement time
limit.
Check cuff and hose.
Repeat NIBP measurement.
REPLACE BATTERY Battery voltage is less than absolute minimum. Shutdown
imminent.
Replace with charged battery.
REPORT FULL Summary report memory full.
Erase summary report.
RETRY ANALYSIS Advisory message in conjunction with noisy ECG. Analysis
halted.
RPT NO DISK Internal disk is not formatted and cannot be formatted.
SELECT 30J FOR TEST Attempt to run a self test at an energy other than 30J.
SELECT ASYNC PACE User is prompted to select Async pacing, since the PD
module cannot detect sync pulses.
SELECT DEFIB MODE Analyze button pressed in pace or monitor mode.
Turn main selector knob to Defib/ON.
SELECT LIMB LEADS Paddles or augmented ECG leads selected when
continuous analysis active or started.
Select limb leads I, II, III or MFE.
SELECT PADS ECG lead (non-MFE Pads) selected when ANALYZE
pressed.
SET CLOCK Real time clock failure: invalid date or time.
Set time and date information.
Verify that the internal lithium battery has been
replaced within the last 5 years. Contact ZOLL
Technical Service Department for assistance.
SET PACE MA Multiple copy errors are the product of intended software
or memory errors. If error reoccurs other than on entering
pace the first time or after more than 10 minutes in other
mode, the unit could be broken.
Set pace current. If broken, replace system board.
Error Message Explanation User
Advisory
Technical Action

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB