R Series Service Manual
2-12 9650-0903-01 Rev. L
SHOCK ADVISED Advisory message when Analysis finds a shockable
rhythm. Followed by PRESS SHOCK in Manual Advisory
Defib with auto-charge enabled or in Auto Defib mode, or
by PRESS CHARGE in Manual Advisory Defib with auto-
charge disabled.
SPO2 AMBIENT LIGHT Ambient light is too bright.
Shield sensor from ambient light.
Replace sensor.
Replace SpO
2
module
SPO2 COMM ERROR No transmissions from SpO
2
unit received.
Communication error or no communication from SpO
2
module.
Replace SpO
2
module and/or system board.
SPO2 PULSE SEARCH Pulse search in progress.
STAND CLEAR (Manual Advisory Defib with auto-charge enabled or Semi-
Auto Mode Defib) User pressed Analyze or an analysis
was started automatically as part of the rescue protocol.
Patient rhythm is being analyzed.
SUMMARY TIMEOUT Summary report timeout error.
SYNC DEFIB DISABLED Sync mode active when analyze pressed in defib.
SYSTEM FAULT 210 System task have not been activated for 500 ms – ECG
control.
Replace digital board.
SYSTEM FAULT 211 System task have not been activated for 500 ms – Defib. Replace digital board.
SYSTEM FAULT 212 System task have not been activated for 500 ms – Pace. Replace digital board.
SYSTEM FAULT 212 System task have not been activated for 500 ms – User
Interface.
Replace digital board.
SYSTEM FAULT 214 System task have not been activated for 500 ms – Display. Replace digital board.
SYSTEM FAULT 215 System task have not been activated for 500 ms – Not
used.
Replace digital board.
Error Message Explanation User
Advisory
Technical Action

Table of Contents

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes
Data StorageInternal memory
ConnectivityUSB

Related product manuals