6253 DC Voltage Current Source/Monitor Operation Manual
5.8.3 Remote Command List
5-34
Source Trigger mode M0 AUTO
Ye s Yes
M1 HOLD
M? Response: M0 or M1
Yes Yes
ST0 AUTO
Ye s Yes
ST1 HOLD
• This command has the same function as the M
command.
ST? Response: STS0 or ST1
Yes Yes
Operate/
Standby
SBY Sets the output to OFF (Standby).
Ye s Ye s
OPR Sets the output to ON (Operate).
SUS Sets the output to Suspend
SBY?,
OPR?,
SUS?
The present output status is returned.
Response:
Yes Yes
Remote
sensing
RS0 2 W
Ye s
RS1 4 W
RS? Response: RS0 or RS1
Yes Yes
Time
parameter
SP
<Th>,
<Td>, <Tp>
[, <Tw>]
Th : Hold time
0 ms Yes
Td : Measurement delay time Unit: ms
4 ms
Tp : Period Tw can be omitted.
50 ms
Tw : Pulse width
25 ms
SP? Response: SP<Th>, <Td>, <Tp>, <Tw>
Th, Td, Tp, Tw: <d.dddd>
*1
Yes Yes
SD Tds Tds : Source delay time (unit: ms)
0.005 ms Yes
SD? Response: SDd.dddd
*1
Yes Yes
Response FL0 SLOW
Ye s
FL1 FAST
FL? Response: FL0 or FL1
Yes Yes
Slew rate SRS0 OFF
No No
SRS1 ON
SRS? Response: SRS0 or SRS1
Yes Yes
*1: The decimal point position in response varies depending on the setting value.
For the source value, limit value, and time parameter setting ranges, refer to the specifications.
Item Command Description
Default Operation possibility
Power
ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS
Status Response
Operate OPR
Suspend SUS
Standby SBY