6253 DC Voltage Current Source/Monitor Operation Manual
8.1 Source and Measurement
8-6
Switching noise
*4) The limit is inactive. While the limit is active, it is the same as the current source range switching
noise.
*5) “digits” indicates current source 5
1
/
2
digit values.
Settling time:
Time to reach the final value 0.1% at pure resistance load and load capacity of 2.5 pF or less and
with the compliance set to the full-scale value
Voltage source in the 100 V range: at load where the output current falls within 20% or less of the
setting limit value
Typical value [p-p] Load resistance
Output ON/OFF noise Voltage source 600 mV At 100 k
Current source 600 mV At 100 k
Range switching noise Voltage source 50 mV
Voltage limit 50 mV
*4)
-
Voltage measurement 50 mV
*4)
Current source
Current measurement 700 digits + 50 mV
*5)
-
Current limit
Power OFF noise 600 mV At 100 k
(Typical value) Source range Limit range
Output response
FAST SLOW
Voltage source 300 mV
2 A
100 s1 ms3 V
10 V
30 V 200 s2 ms
100 V 300 mA 300 s3 ms
Current source 3 A
100 V
10 ms
30 A5 ms
300 A
2 ms
5 ms
3 mA
30 mA
800 s300 mA
2 A 30 V

Table of Contents