3-40 Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Functions
Trigger Function
Setup and restrictions
• You cannot perform trigger outputs together with trigger inputs. You must select
either trigger output or trigger input.
• To use a trigger function, you must enable the trigger function and select either
TRIG OUT or TRIG IN in the TRIGGER SETUP table on the MEASURE:
OUTPUT SEQUENCE screen. Then the trigger inputs or outputs are performed
automatically after you start a measurement by selecting a measurement
front-panel key (Single, Repeat, or Append).
• The trigger output function is not available for sampling measurements.
• When you perform knob sweep measurements, the trigger function is not
available.
• For the electrical specifications of trigger signals, refer to User's Guide General
Information.