EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>4155B

Agilent Technologies 4155B User Manual

Agilent Technologies 4155B
414 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #186 background imageLoading...
Page #186 background image
4-44 Agilent 4155B/4156B Users Guide Vol.2, Edition 5
Making a Measurement
Sampling Measurements
To Set up Sampling Parameters
1. Confirm that SAMPLING is set in the MEASUREMENT MODE field on the
CHANNELS: CHANNEL DEFINITION screen. If SAMPLING is not set, select
SAMPLING secondary softkey in the MEASUREMENT MODE field.
2. Press Meas key in the PAGE CONTROL key group.
3. Select SAMPLING SETUP primary softkey.
4. In the MODE field of SAMPLING PARAMETER, select:
LINEAR secondary softkey for equally spaced sampling intervals.
LOG XX secondary softkey for logarithmically spaced sampling intervals.
XX is 10, 25, or 50 sampling points per decade.
THINNED-OUT secondary softkey for reduced sampling interval of more
recent samples (by thinning less recent samples).
For details about sampling mode, see Sampling Measurement Mode in
Chapter 2.
5. In the INITIAL INTERVAL field, enter a value for the first sampling interval.
6. In the NO. OF SAMPLES field, enter the number of points at which to sample.
7. If you select LINEAR or THINNED-OUT in MODE field, set the TOTAL
SAMP. TIME (total sampling time) which specifies the time from the start of
sampling to the end. This field is not available for the logarithmic sampling.
In the TOTAL SAMP. TIME field, enter a value or select:
NO LIMIT secondary softkey for excluding the total sampling time from the
sampling completion conditions.
AUTO secondary softkey for excluding the total sampling time from the
sampling completion conditions, and including the number of samples to the
completion conditions. This softkey is available only for the linear sampling.
For details about sampling completion conditions, see To Define Stop
Conditions on page 4-47.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155B and is the answer not in the manual?

Agilent Technologies 4155B Specifications

General IconGeneral
BrandAgilent Technologies
Model4155B
CategoryMeasuring Instruments
LanguageEnglish