4-28 Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Making a Measurement
Sweep Measurements
To Display Graphics Results
1. Press Display key in the PAGE CONTROL key group.
2. Select DISPLAY SETUP primary softkey.
3. In the DISPLAY MODE field, select GRAPHICS secondary softkey.
4. In the X axis column, enter variable name, select axis scale, and enter minimum
and maximum values.
5. In the Y1 axis column, enter variable name, select axis scale, and enter
minimum and maximum values.
6. If you use Y2 axis, enter variable name, select axis scale, and enter minimum
and maximum values in Y2 axis column.
When the pointer is in the NAME row, the allowable variable names appear in the
secondary softkey area. To set a variable name, select the desired secondary softkey.
The allowable names are names that you already set up on the CHANNEL
DEFINITION, USER FUNCTION, and USER VARIABLE screens.
To display a grid on the plotting area
In GRID field, select ON secondary softkey.
To remove the grid
In GRID field, select OFF secondary softkey.
To control display of line parameters on GRAPHICS screen
In LINE PARAMETER field, select ON to display or OFF to not display. Line
parameters are the X and Y intercepts and gradient of the analysis lines.
To set up variable to be displayed on the GRAPHICS screen
In DATA VARIABLES fields, select secondary softkey for desired variable.