4-60 Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Making a Measurement
Stress Force
To Use Selector
1. Press Stress key in the PAGE CONTROL key group.
2. Select CHANNEL DEF primary softkey.
3. In the MEASURE field of the desired channel in the SMU/PG SELECTOR area,
select:
• SMU secondary softkey to connect SMU to DUT during measurement state.
• PGU secondary softkey to connect PGU to DUT during measurement state.
• OPEN secondary softkey to disconnect SMU, PGU, and DUT during
measurement state.
• PGU OPEN secondary softkey to disconnect PGU from DUT by
semiconductor switch during measurement state.
4. In the STRESS field of the desired channel in the SMU/PG SELECTOR area,
select:
• SMU secondary softkey to connect SMU to DUT during stress force state.
• PGU secondary softkey to connect PGU to DUT during stress force state.
• OPEN secondary softkey to disconnect SMU, PGU, and DUT during stress
force state.
• PGU OPEN secondary softkey to disconnect PGU from DUT by
semiconductor switch during stress force state.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155B and is the answer not in the manual?

Agilent Technologies 4155B Specifications

General IconGeneral
BrandAgilent Technologies
Model4155B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals