Agilent 4155B/4156B Users Guide Vol.2, Edition 5 4-21
Making a Measurement
Sweep Measurements
To Set up SMU Pulsed Output
SMU pulse output source is the measurement unit defined as VPULSE or IPULSE
in the CHANNELS: CHANNEL DEFINITION screen. For pulsed sweep source, set
the function (FCTN) to VAR1, VAR2, or VAR1. For pulsed constant source, set the
function to CONST. To set up the SMU pulse output source, press Meas front-panel
key. The MEASURE: SWEEP SETUP screen is displayed.
1. Set the source output parameters:
For VAR1, refer to To Set up Primary Sweep on page 4-14.
For VAR2, refer to To Set up Secondary Sweep on page 4-16.
For VAR1, refer to To Set up Synchronous Sweep on page 4-18.
For CONST, refer to To Set up Constant Output on page 4-20.
2. SMU PULSE: PERIOD
Enter the pulse period value.
3. SMU PULSE: WIDTH
Enter the pulse width value.
4. SMU PULSE: BASE
Enter the pulse base value.
You cannot change UNIT and NAME in this screen. To change the values, go to the
CHANNELS: CHANNEL DEFINITION screen.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155B and is the answer not in the manual?

Agilent Technologies 4155B Specifications

General IconGeneral
BrandAgilent Technologies
Model4155B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals