3-20 Agilent 4155B/4156B Users Guide Vol.2, Edition 5
Measurement Functions
Knob Sweep Function
VAR1 SETUP Softkey
This softkey displays secondary softkeys for setting the primary sweep source
(VAR1) parameters.
SWEEP
MODE
SINGLE Selects sweep mode. This softkey displays present setting.
setting
Pressing this softkey toggles the sweep mode in the following
order:
SINGLE ® DOUBLE ® SINGLE
default
setting of the SWEEP MODE field on the MEASURE: SWEEP
SETUP screen
POLAR-
ITY
POS Selects polarity of sweep source. Changing the setting resets the sweep
step to 0, so the sweep measurement curve goes back to 0 on the screen.
For details, see Sweep Step Value on page 3-15. This softkey
displays present setting.
setting
Pressing this softkey toggles the polarity of VAR1 channel in the
following order:
POS ® NEG ® BIPOLAR ® POS
default
If polarity of VAR1 stop and start value on MEASURE: SWEEP
SETUP screen are same, default value is same as polarity of stop
and start value.
If polarity of VAR1 stop and start value on MEASURE: SWEEP
SETUP screen are different, default value is BIPOLAR.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155B and is the answer not in the manual?

Agilent Technologies 4155B Specifications

General IconGeneral
BrandAgilent Technologies
Model4155B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals