Agilent 4155B/4156B Users Guide Vol.2, Edition 5 4-33
Making a Measurement
Sweep Measurements
To Use R-Box
1. Connect the 16441A R-Box to the 4155B/4156B and to the 16442A Test Fixture
or connector plate on your shield box. For details about connections, refer to
R-BOX Control in Chapter 3.
2. Press Chan front-panel key of the PAGE CONTROL key group.
3. Select CHANNEL DEF primary softkey to display the CHANNELS:
CHANNEL DEFINITION screen.
4. In the SERIES RESISTANCE fields, select:
0 ohm secondary softkey to connect 0 W resistance.
10k ohm secondary softkey to connect 10 kW resistance.
100k ohm secondary softkey to connect 100 kW resistance.
1M ohm secondary softkey to connect 1 MW resistance.
Resistance is switched just before and just after measurement state. In the standby
state, the stress force state, and the idle state, 0 W is connected.
The 4155B/4156B automatically compensates for the resistance values.
For the following SMUs, you can set 0 W only:
SMU that is set to ON in the STBY field.
SMU that is set to COMMON in the MODE field.
You can set resistance values for the following SMUs.
If the 41501A/B SMU/Pulse Generator Expander is not installed or does not
have an HPSMU
SMU1
SMU2
If the 41501A/B has an HPSMU
SMU1
SMU5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155B and is the answer not in the manual?

Agilent Technologies 4155B Specifications

General IconGeneral
BrandAgilent Technologies
Model4155B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals